Sedaconda ACD-L
Sedaconda ACD-L
Sedaconda ACD-L is a medical device which enables the administration of volatile anaesthetics (isoflurane or sevoflurane) and is inserted between the Y-piece and ET-tube. The device has no electrical components and is MRI & CT scan compatible. It also acts as an efficient Heat and Moisture Exchanger (HME) and contains an electrostatic bacterial/viral filter. The Sedaconda ACD-L has a dead space of 100 ml and can function with tidal volumes as low as 350 ml.
- Use: Single Patient, 24 hour use
- Device Classification: Class IIa
- Shelf Life: 5 Years from Date of Manufacturing
- Reference Number: 1026100
- Packaging unit: 1 unit contains 12 Sedaconda ACD-L devices
Operating Conditions
Disposable product: |
Single patient, 24 hours use |
Anaesthetic agents: |
isoflurane or sevoflurane |
Efficiency: |
Normally >90% is reflected. Depending on Minute Volume. |
Tidal volume working range: |
Minimum 350 ml |
Resistance to gas flow @ 60 L/min: |
2.5 cm H₂O (250 Pa) |
Moisture loss at 0.75 l x 12 breaths/min: |
5 mg/l |
Corresponding moisture output: |
30 mg H₂O/l |
Moisture loss at 1.0 l x 10 breaths/min: |
7 mg/l |
Corresponding moisture output: |
30 mg H₂O/l |
Bacterial filtration: |
99.867% |
Viral filtration: |
99.76% |
Dead space: |
Approx. 100 ml |
Connectors: |
15M/22M-15F |
Weight: |
50 g |
Gas sampling port: |
Female Luer lock |
Agent supply line: |
2200 mm |
Instruction for Use, Standard placement (IFU) Rev.2.5 2022-03