Sedaconda ACD-L
Sedaconda ACD-L
Sedaconda ACD-L is a disposable medical device which enables the administration of volatile anaesthetics (isoflurane or sevoflurane) and is inserted between the Y-piece and ET-tube. The device has no electrical components and is MRI & CT scan compatible. It also acts as an efficient Heat and Moisture Exchanger (HME) and contains an electrostatic bacterial/viral filter. The Sedaconda ACD-L has a dead space of 100 ml and can function with tidal volumes as low as 350 ml.
- Use: Single Patient, 24 hour use
- Device Classification: Class IIa
- Shelf Life: 5 Years from Date of Manufacturing
- Reference Number: 1026100
- Packaging unit: 1 unit contains 12 Sedaconda ACD-L devices
Operating Conditions
Disposable product: |
Single patient, 24 hours use |
Anaesthetic agents: |
isoflurane or sevoflurane |
Efficiency: |
Normally >90% is reflected. Depending on Minute Volume. |
Tidal volume working range: |
Minimum 350 ml |
Resistance to gas flow @ 60 L/min: |
2.5 cm H₂O (250 Pa) |
Moisture loss at 0.75 l x 12 breaths/min: |
5 mg/l |
Corresponding moisture output: |
30 mg H₂O/l |
Moisture loss at 1.0 l x 10 breaths/min: |
7 mg/l |
Corresponding moisture output: |
30 mg H₂O/l |
Bacterial filtration: |
99.867% |
Viral filtration: |
99.76% |
Dead space: |
Approx. 100 ml |
Connectors: |
15M/22M-15F |
Weight: |
50 g |
Gas sampling port: |
Female Luer lock |
Agent supply line: |
2200 mm |
Instruction for Use (IFU), Rev 5
Previous versions
Instruction for Use (IFU), Rev 4
Instruction for Use, Standard placement (IFU) Rev.3
Instruction for Use, Alternativ placement (IFU). Rev.3
Instruction for Use, Standard placement (IFU) Rev.2.5
Instruction for Use, Alternativ placement (IFU). Rev.2.5
Instruction for Use, Standard placement (IFU). Rev.2
Instruction for Use, Alternativ placement (IFU). Rev.2