Sedaconda ACD-S is a medical device which enables the administration of volatile anaesthetics (isoflurane or sevoflurane) and is inserted between the Y-piece and ET-tube. The device has no electrical components and is MRI & CT scan compatible. It also acts as an efficient Heat and Moisture Exchanger (HME) and contains an electrostatic bacterial/viral filter. The Sedaconda ACD-S has a dead space of 50 ml and can function with tidal volumes as low as 200 ml.
| Disposable product: | Single patient, 24 hours use |
| Anaesthetic agents: | isoflurane or sevoflurane |
| Efficiency: | Normally 88% is reflected. Depending on Minute Volume. |
| Tidal volume working range: | Minimum 200 ml |
| Moisture loss at 0.5 l x 12 breaths/min: | 5 mg/l |
| Corresponding moisture output: | 30 mg H₂O/l |
| Moisture loss at 0.7 l x 10 breaths/min: | 6 mg/l |
| Corresponding moisture output: | 29 mg H₂O/l |
| Bacterial filtration: | 99.867% |
| Viral filtration: | 99.76% |
| Dead space: | Approx. 50 ml |
| Connectors: | 15M/22M-15F |
| Weight: | 38 g |
| Gas sampling port: | Female Luer lock |
| Agent supply line: | 2200 mm |